Model
|
SADP16
|
Cable
|
Flexible towline cable
|
Structure
|
Plunger
|
Liquid Level Detection
|
pLLD
|
Function
|
1.Automatic pick-up, eject the tips(require Z-axis)
2.Tip presence sensor
3.Liquid level detection
4.Liquid level following(require Keyto Z-axis)
5.Clot detection
6.Anti-droplet control
7.Bubbles/foam detection
8.Aspiration of air detection
|
Pipetting Range
|
1ul-1ml
|
Fastest Aspirete and Dispense Speed
|
2000ul/s
|
Fastest Liquid Level Detection Speed
|
10ms
|
CV at Minimum Dispense Volume @50ul tip
|
1ul, 5%
|
CV%, Dispense 5 ul @50ul tip
|
2%
|
CV%, Dispense 10 ul @200ul tip
|
2%
|
CV%, Aliquot Dispense 20 ul @1000 ul tip
|
3%
|
Tip Volume
|
10μL/50μL/200μL/1000μL
|
Minimum Tip-to-tip Spacing
|
9mm
|
Communication
|
RS-232, RS-485, CAN
|
Motor
|
Coreless brushless motor
|
Z-axis Compatible
|
Z180
|
Movement Mechanism Compatible
|
Y
|
Aspiration Monitor
|
Y
|
Lifetime
|
1 Million Cycles
|
Power
|
24VDC
|
Max Pipetting Units
|
8 units
|
Dimensions
|
265*100*16 mm
|
Weight
|
≤330g
|